PM2.5 or PM10 Ambient Air Monitoring With TSM
Made in the USA
PM2.5 or PM10 Ambient Air Monitoring With TSM This unit is the most advanced particulate collection system in the industry. This can be configured for TSP, PM10, PM2.5, PM1 as well as any other particulate separation device with a flow rate between 1-25 SLPM. Featuring modern, industrially hardened components paired with innovative design and state of the art manufacturing practices, the TE-Wilbur is the future of particulate collection. The unit is ideal for Environmental Monitoring, Fence Line Monitoring and Industrial Monitoring. This is designed to operate in any environment and is set up for quick and easy deployment. The TE-Wilbur exceeds all performance specifications set forth by the USEPA for low-volume particulate collection.
Exceeds USEPA 40 cfr part 53, 50 appendix.
5.7 inch color touch screen with intuitive control Multi-language Support Data retrieval via USB and MODBUS Preprogrammed for USEPA sample schedules Integrated Uninterrupted power supply Low maintenance / cost to operate Brushless DC pump.
This most advanced particulate collection system in the industry. The unit can be configured for TSP, PM10, PM2.5, PM1 as well as any other particulate separation device with a flow rate between 1-25 SLPM.Featuring modern, industrially hardened components paired with innovative design and state of the art manufacturing practices, the TE-Wilbur is the future of particulate collection. This is ideal for Environmental Monitoring, Fence Line Monitoring and Industrial Monitoring. This unit is designed to operate in any environment and is set up for quick and easy deployment. This unit exceeds all performance specifications set forth by the USEPA for low-volume particulate collection. This is user interface is a full-color 5.7" touchscreen which utilizes intuitive control. The touchscreen was developed specifically to satisfy the recommendations of instrumentation operators. This features easy to access menus, step by step calibration and on screen status alerts. The unit control is provided by our industrialized programmable logic controller (PLC) whose programming is USB downloadable and MODBUS compatible. Volumetric flow is controlled through an integrated mass flow meter and onboard ambient temperature and pressure sensors. Filter set-up and exchange is achieved through the designed filter exchange mechanism which allows easy one handed operation. In addition, the filter exchange mechanism does not require continual adjustment to seal your filter against leaks. Data can be retrieved through any MODBUS compatible data acquisition system or can be easily downloaded and transported with your filter on a USB drive. Industrial PLC control Integrated step by step calibration.
Particulate Size:
TSP, PM10, PM2.5, PM1, and any size selectable inlet which operates between 1-25 slpm.
Sampling Interval :
Preprogrammed USEPA 1 in 3, 1 in 6, 1 in 12, 24hr sampling User defined sampling intervals Meteorological sampling triggers.
Operating Temperature: -25°C to 50°C
Ambient Temperature Sensor : Sensor: PT100 RTD, Accuracy: ±0.15°C , Resolution: 0.01°C
Filter Temperature Sensor: Sensor: PT100 RTD, Accuracy: ±0.15°C, Resolution: 0.01°C
Barometric Pressure Sensor: Range: 450 mmHg to 1238 mmHg
Accuracy: ±10.00 mmHg, Resolution: 0.75 mmHg
Flow Control Specification: Brushless DC Pump : Range: 0-25 Liters per minute
Accuracy: ±2% or less at 16.67 LPM ; Repeatability: ±1% of full scale (0.25 LPM) ; Resolution: 0.01 LPM
Power Requirements : 120V/240V, 50/60Hz
Power Consumption : 40 watt (peak)
Weight : 48lbs / 21.8kg without stand
Dimensions : 20”W x 20”H x 10”D OR 508mm x 508mm x 254mm